FEATURES
• -35 to +80°C (-31 to +176°F) measurement range
• USB interface for set-up and data download
• User-programmable alarm thresholds
• Status indication via red, green and orange LEDs
• Supplied with replaceable internal lithium battery and Windows control software
• Environmental protection to IP67
|
Specifications |
Minimum |
Typical |
Maximum |
Unit |
|
Measurement range |
-35 (31) |
|
+80 (+176) |
°C (°F) |
|
Internal resolution |
|
0.5 (1) |
|
°C (°F) |
|
Accuracy (overall error) |
|
±1 (±2) |
|
°C (°F) |
|
Logging rate |
every 10s |
|
every 12hr |
- |
|
Operating temperature range |
-35 (-31) |
|
+80 (+176) |
°C (°F) |
|
1/2AA 3.6V Lithium Battery Life |
|
1* |
|
Year |
* At 25°C and 1 minute logging rate
This standalone data logger measures and stores up to 16,382 temperature readings over a -35 to +80°C (-31 to +176°F) range.
The user can easily set up the logging rate and start-time, and download the stored data by plugging the data logger into a PC’s USB port and running the purpose designed software under Windows 2000, XP and Vista (32-bit). Data can then be graphed, printed and exported to other applications. The data logger is supplied complete with a long-life lithium battery which can typically allow logging for 1 year. Correct functioning of the unit is indicated by flashing red, green and orange LEDs. The data logger is protected against moisture to IP67 standard when the protective cap is fitted.
EL-WIN-USB (CONTROL SOFTWARE)
The software allows the following parameters to be configured:
- Logger name
- °C, °F
- Logging rate (10s, 1m, 5m, 30m, 1hr, 6hr, 12hr)
- High and low alarms
- Immediate, delayed and push-to-start logging
- Display off, on for 30 seconds after button press, or permanently on
- Data rollover (allows unlimited logging periods by overwriting the oldest data when the memory is full)
|
|
Green double flash
The data logger is not currently logging, but is primed to start at a later date and time (delayed start). |
|
Green single flash
The data logger is currently logging. No alarm. |
|
|
Green and Orange single flash
The data logger is full and has stopped logging. No alarm. |
|
|
Red single flash
The data logger is currently logging. Low alarm. |
|
|
Red and Orange single flash
The data logger is full and has stopped logging. Low alarm. |
|
|
Red double flash
The data logger is currently logging. High alarm. |
|
|
Red and Orange double flash
The data logger is full and has stopped logging. High alarm. |
|
|
Orange single flash
The battery is running low as its voltage has dropped below 3.3V. Data logging will continue until the battery voltage drops below 2.8V. |
|
|
No LEDs flash
The data logger is stopped, the battery is empty or there is no battery. |